The Ganymede Series SD-OCT is a highly customizable Optical Coherence Tomography system with the ability of producing high-resolution 2D cross-sectional and 3D volumetric images of a sample. The system is capable of A-scan line rates up to 248 kHz along with 64-bit software pre-installed on the included computer which displays and processes 2D and 3D OCT data in real time.
SD-OCT Systems with 880 nm Center Wavelength (See Specifications Tab)
A-Scan Rates up to 248 kHz
Imaging Depths in Air up to 3.4 mm
Axial Resolutions in Air from <3.0 to 6.0 µm
Sensitivities up to 106 dB
Includes Computer and ThorImage OCT Software Package
Build-Your-Own and Preconfigured Systems Available
Thorlabs’ Ganymede™ Series of OCT Imaging Systems provide the flexibility required for high-resolution imaging applications with A-scan line rates up to 248 kHz. The 64-bit software pre-installed on the included computer displays and processes 2D and 3D OCT data in real time. Scanner options include a robust standard scanner and a user-customizable scanner. Optional accessories are available to customize your OCT system to meet the requirements of your application. Additionally, we offers five complete, preconfigured OCT systems operating with 880 nm centre wavelengths.
Imperial Systems: OCT-STAND (Stand) and OCT-XYR1 (Stage)
Metric Systems: OCT-STAND/M (Stand) and OCT-XYR1/M (Stage)
Specifications
Preconfigured System Key Specifications
System Item #
GAN111C1(/M)|
GAN312C1(/M)|
GAN332C1(/M)|
GAN612C1(/M)|
GAN632C1(/M)|
Centre Wavelength
880 nm
Imaging Depth (Air/Water)
3.4 mm / 2.5 mm
3.4 mm / 2.5 mm
1.6 mm / 1.2 mm
3.4 mm / 2.5 mm
1.6 mm / 1.2 mm
Axial Resolution (Air/Water)
6.0 µm / 4.5 µm
6.0 µm / 4.5 µm
<3.0 µm / <2.2 µm
6.0 µm / 4.5 µm
<3.0 µm / <2.2 µm
Lateral Resolution
8.0 µm
8.0 µm
4.0 µm
8.0 µm
4.0 µm
A-Scan/Line Rate
1.5 – 20 kHza
1.5 – 80 kHzb
5 – 248 kHzc
Sensitivity (Max)d
106 dB (at 1.5 kHz)
102 dB (at 5 kHz)
Software
ThorImageOCT Software
Interactive Scan Position Control through Video Display for Common Line Scans or Freeform Pattern Scans
Advanced Dataset Management
Access to Raw Spectra, Processsed Data, and All Calibration Files Necessary for User-Designed Processing Routines
High-Speed Volume Rendering of 3D Data
Doppler and Speckle Variance Imaging
Versatile Scan and Acquisition Control, such as Averaging or Adjustable Scan Speeds
ThorImageOCT is a high-performance data acquisition software that is included with all Thorlabs OCT systems. This 64-bit Windows-based software acquires and displays OCT data, as well as includes scan control and processing options. Additionally, NI LabVIEW and C-based Software Development Kits (SDKs) are available, which contain a complete set of libraries for measurement control, data acquisition, and processing, as well as for storage and display of OCT images. The SDKs provide the means for developing highly specialized OCT imaging software for every individual application.